USB-MCA [Applicable Models] APG7300A |
Q1 | Does this product have a list mode function? |
A1 | No there is not. APG7400A USB-MCA4 which is a high - end model has its function. |
Top |
Q2 | "Connection error" dialog box is displayed when starting the attached application, and the computer and the device can not be connected. |
A2 | Please refer to "Ethernet". |
Top |
Q3 | Does it support USB 3.0? |
A3 | Connection and use with the USB 3.0 port are available. However, the communication speed is USB 2.0 specification. |
Top |
Q4 | Do you offer sample programs? |
A4 | No there is not. APG7400A USB-MCA4 which is a high - end model has its function. Please download from this page and use it. The development language is Visual C ++, Visual C #, Linux, LabVIEW. |
Top |
Q5 | Does this product have a Multi Channel Scaler (MCS) function? |
A5 | No there is not. APG7400A USB-MCA4 which is a high - end model has its function. |
Top |
Q6 | Does this product have a Coincidence function? |
A6 | No there is not. APG7400A USB-MCA4 which is a high - end model has its function. |
Top |
Q7 | Is there a dead time? How much is that? |
A7 | Yes, The dead time is a time period when MCA can not measure peak value. New events can not be measured while events occur irregularly, such as radiation, while events are being detected, peak detection, digitization of wave height values, memory rewriting, resetting of crest value. The dead time of this device is fixed 500 ns. |
Top |
Q8 | Does this product have a GATE function? |
A8 | Yes. To acquire event data at that time according to external conditions when an event occurs, input the LV - TTL level signal to the "GATE" LEMO connector on the front panel. It is measured when it is High, but it is not measured when it is Low. |
Top |
Q9 | Does this product have a VETO function? |
A9 | Yes. If you want to discard event data at that time due to external conditions at the occurrence of an event, input the LV - TTL level signal to the "VETO" LEMO connector on the front panel. This has the opposite effect to "GATE". Measurement is performed when it is Low, but not when it is Low. The timing is the same as GATE mentioned above. |
Top |
Q10 | Can I reset the time information externally during measurement? |
A10 | No, it does not correspond. |
Top |
Q11 | How can I connect the cable with BNC connector from the detector to the USB-MCA? |
A11 | Signal input connector to this device is "EPL.00.250.NTN"
made by LEMO or its equivalent shape. When using a cable with a BNC connector, it is possible to connect with this device using the following conversion adapter. Manufacturer: Huber & Suhner Manufacturer model: 33_QLA-BNC-01-1/1-_NE Contents: QLA-01 to BNC Connector Gender 1 Interface: QLA-01 Connector Gender 2 Interface: BNC |
Top |
Q12 | How do I install the driver software? |
A12 | Please refer to the "Installing driver software" chapter in the attached user manual. Or please refer to the latest user manual from our website. |
Top |