Support > Question and Answer > DSP for Gamma-ray

 Digital Signal Processor

 [Applicable Models]
 APN (U) 101

Q1  "Connection error" dialog box is displayed when starting the attached application, and the computer and the device can not be connected.
Q2  Do you have the recommended settings?
Q3  Do you offer sample programs?
Q4  It seems that energy resolution is bad...
Q5  It seems that the count rate is low...

Q1  "Connection error" dialog box is displayed when starting the attached application, and the computer and the device can not be connected. 
A1  Please refer to "Ethernet".
Top
Q2  Do you have the recommended settings?
A2  Please refer to "Configuration Reference".
 NOTE: The content of setting varies depending on the type and polarity of the detector, importance on energy resolution capability or emphasis on counting.
Top
Q3  Do you offer sample programs
A3  For now we offer sample programs only for a few products. Please download from this page and use it.. The development language is Visual C ++, Visual C #, Linux, LabVIEW.
Top
Q4  It seems that energy resolution is bad...
A4  As an example of the DSP for gamma ray measurement, we confirmed that it has the performance shown in the figure below in relation to the resolution of 1.33 MeV of Co - 60 and the counting rate.

ICRvsFWHM

 The higher the count rate, the worse the energy resolution becomes. If Risetime is lengthened to 6 ƒÊs as shown above, good energy resolution can be obtained when the count rate is low. As the counting rate increases, good energy resolution can not be maintained. On the other hand, if Risetime is shortened to 0.5 ƒÊs, the energy resolution will be low even when the counting rate is low, but its energy resolution will be maintained to some extent even if the counting rate increases.

 If the energy resolution is bad, please check the following.
Please refer to the hardware manual and software manual for details.

1. The power supply of the DSP is taken from the same tap supplying the preamplifier power of the detector.

2. There are no equipment near the DSP that will become a noise generation source like a refrigerator or inverter.

3. Do not bring the preamplifier output signal cable of the detector close to the power cable of AC100V. Also, there is no equipment such as a computer power adapter near the cable as a noise source.

4. The setting of the gain of the preamplifier output signal captured inside the DSP is appropriate. Connect the "MONI" output terminal to the oscilloscope with a cable, set the DAC type of the attached application as "preamp", the signal is within the upper limit 1 V and the influence of noise is minimal. Also, the pole zero is properly adjusted.

5. "Analog trigger threshold" setting is appropriate. Start measurement and find the boundary with the noise level while looking at the value of "input total rate". In the case of the noise level, the value becomes extremely large. If you raise the set value gradually, the noise will not come out. Set a value that is slightly higher than the noise level setting.

6. "Slow trigger threshold" setting is appropriate. Start measurement and confirm that "input total rate" is not extremely high count due to noise influence. Next, look at the value of "throughput rate" and find the boundary with the noise level. In the case of the noise level, the value becomes extremely large. If you raise the set value gradually, the noise will not come out. Set a value that is slightly higher than the noise level setting.

7. The waveform shaping (filtering) signal inside the DSP is appropriate. Connect the "MONI" output terminal to the oscilloscope with a cable, set the DAC type of the attached application to "slow", and that signal is properly adjusted within the range of 1 V upper limit.

8. If there are variations in the rising edge of the preamplifier output waveform, such as germanium semiconductor detector, set the value of "slow flattop time" to twice the rise time considered to be the slowest.

9. High voltage power supply to the detector is not appropriate. Also check the rated voltage, ripple, cable and so on. .

10. Power supply to the preamplifier power supply is not appropriate.
Top
Q5  It seems that the count rate is low...
A5  As an example of the DSP for measuring gamma rays, we confirmed that it has the performance as shown below in the relation between count rate input total rate (cps) and throughput rate (cps).

ICRvsOCR

If the above performance is not good, please check the following.

1. "Analog coarse gain" and "analog fine gain" setting value or front panel "F. G" volume setting is high.

2. The setting value of "fast trigger the shold" is high.

3. "Slow trigger theshold" setting value is high.

4. The setting value of "slow risetime" is large.

5. "Slow flattop time" setting value is large. Set with the minimum value that can absorb the rising edge of the preamplifier output signal.
Top